IEC 60068 Desktop Constant Temperature And Humidity Test Chamber – BONAD Environmental Test Equipment
Standards:
This test chamber complies with key international standards:
- IEC 60068-2-1: Environmental testing for electrical and electronic products – Low temperature tests.
- IEC 60068-2-2: High temperature tests.
- IEC 60068-2-78: Constant humidity and heat tests.
- EN 14511
- ASTM E1171
Applications:
The chamber is essential for quality inspection across various product categories such as electronics, electrical components, communication devices, instruments and meters, vehicles, plastics, metals, food products, chemicals, building materials, medical devices, and aerospace components.
Features:
- Precise Control: Utilizes high-precision sensors and advanced PID control algorithms to maintain uniform temperature and humidity with minimal fluctuations.
- Wide Range: Supports temperatures from -70℃ to +150℃ and humidity levels from 20% to 98% RH to accommodate diverse testing needs.
- Rapid Temperature Changes: Equipped with efficient refrigeration and heating systems for quick temperature adjustments to enhance test efficiency.
- User-Friendly Interface: Features a large touch screen interface for easy operation; displays real-time data including temperature/humidity curves and fault alarms.
- Safety Protections: Incorporates multiple safety features like over-temperature protection, water shortage alerts, and overload safeguards ensuring reliable operation.
- Energy Efficiency: Employs eco-friendly refrigerants alongside energy-saving designs to minimize operational costs while meeting environmental standards.
Testing Capabilities:
The benchtop test chamber supports various testing scenarios:
- High Temperature & Humidity Tests: Assess product stability under extreme conditions.
- Low Temperature & Humidity Tests: Evaluate performance in cold environments.
- Temperature & Humidity Cycles: Detect potential defects through rapid environmental changes.
- Constant Storage Conditions: Provide stable environments for sensitive materials or products.
Technical parameters:
Parameter | Specification |
---|---|
Volume (L) | 225 L |
Low Temperature Range | F: RT +10℃ / Z: 0℃ / R: -20℃ / L: -40℃ / U: -60℃ / S: -70℃ (Customizable) |
High Temperature Range | +150℃ (180℃ / 200℃ Special Order) |
Humidity Range | 20% – 98% R.H (5% – 98% R.H; 10% – 98% R.H as special selection conditions) |
Temperature Uniformity | ±1℃ |
Temperature Fluctuation | ±0.5℃ |
Humidity Uniformity | ±3% R.H |
Humidity Fluctuation | ±2% R.H |
Heating Rate | 2.0 – 4.0℃/min (Customizable heating rate as per customer requirements) |
Cooling Rate | 0.7 – 1.0℃/min (Customizable cooling rate as per customer requirements) |